Illustration shows complete Oxygen Outlet Point (with Twin
Pipeline Adaptor & BPC Flowmeters), as would be seen in
actual use in war ds. |
 |
|
| |
3303 - Oxygen 3304 - Nitrous Oxide
3305 - Air 3306 - V acuum
The outlet points from which gases & vacuum are drawn are
of self sealing typ e and hav e non-interchangeable outlet
fittings. The supply is established simply by inserting the
appropriate safet y key plug into the orifice of the outlet
point and pressing it home until it locks in position. To break
the connection, the milled collar round the orifice is rotated
a few degrees to the right, when the safet y key plug is
automatically ejected and the gas flow is cut off. |
 |
|
|
| |
3307 - Oxygen 3308 - Nitrous Oxide
3309 - Air 3310 - V acuum
These are non-interchangeable like the self sealing valves
and thus eliminate the possibility of errors in making
connections. |
 |
|
|
| |
3316
To keep the cov ers in position on the ‘S’ bracket. |
 |
|
|
| |
3315
For connecting the down drop pipes to self sealing
valves. |
 |
|
|
| |
3311 - Oxygen 3312 - Nitrous Oxide
3313 - Air 3314 - V acuum
Coloured coded text for respective self sealing valves to safe
guard against the possibility of administering the wrong
gas. |
 |
|
|
| |
3317 - 10mm OD 3318 - 16mm OD 3319 - 28mm OD
For surface mounting pipes of different sizes, firmly on the wall. |
 |
|
|
| |
3322 - 15mm x 8mm 3323 - 15mm x 15mm |
 |
|
|
| |
3320 - 15 x 15 x 8mm 3321 - 15 x 15 x 15mm |
 |
|
|
| |
3324 - 0 - 250kg/cm2 High Pressure, Oxygen, Dia 100mm
3326 - 0 - 10kg/cm2 Low Pressure, Oxygen, Dia 100mm
3350 - Pipeline Gas Failure Alarm
The contact gauge has electrical contacts. It should be connected to the pipeline gas failure alarm which has two signal
lights, a buzzer & an “ON/OFF” switch. When the contact gauge is connected to the power supply and the ON/OFF switch
of the alarm is switched “ON”, green light will show indicating that the electrical supply serving the contact gauge & alarm
are working correctly. The green light should show at all times when the pipeline is in use. The other signal light viz red is a
warning light which automatically shows & the buzzer rings when the pressure in the running cylinder has fallen to a
determined value. Put the switch to “OFF” position after which the reserve cylinder should be brought into operation
promptly to ensure continuity of supply. |
|
 |
|
| |
 |
|
|
| |
3730 - Grey (Air)
1211 - Yellow (Vacuum)
3729 - Blue (Nitrous Oxide)
3726 - White (Oxygen) |
 |
|